Description
The ERM-II is a radiometer with 3 broadband channels that measure the earth-reflected solar flux and
earth-emitted thermal flux over total (0.2~100μm), long (5~50μm) and short (0.2~5.0μm) wavebands.
Satellite |
FY-3F
|
Band |
Short wavelength |
Long wavelength |
All wavelength |
Spectral range (μm) |
0.2~5.0 |
5.0~50 |
0.2~100 |
Radiance range |
0~280Wm-2Sr-1 |
0~120Wm-2Sr-1 |
0~400Wm-2Sr-1 |
Calibration accuracy |
1% |
0.5% |
0.5% |
Sensitivity |
0.3Wm-2Sr-1 |
0.2Wm-2Sr-1 |
0.2Wm-2Sr-1 |
Characteristics
Channel |
Spectral interval |
Noise Equivalent Radiance |
Absolute accuracy |
Short-wave |
0.2~4.3 µm |
0.4 W⋅m-2⋅sr-1 |
1.5 W⋅m-2⋅sr-1 |
Total radiance |
0.2~50 µm |
0.4 W⋅m-2⋅sr-1 |
1.1 W⋅m-2⋅sr-1 |